キシモト トシユキ   KISHIMOTO TOSHIYUKI
  岸本 敏幸
   所属   朝日大学  歯学部 口腔病態医療学講座 歯科麻酔学
   朝日大学  大学院 歯学研究科
   職種   講師
発表年月日 2018/10/05
発表テーマ The effects of different pulse oximeter application sites on blood oxygen saturation measurements during intravenous sedation
会議名 The 15th International Dental Congress on Anesthesia, Sedation and Pain Control
学会区分 国際学会
発表形式 ポスター
単独共同区分 共同
発表者・共同発表者 Takashi Goto, Takanori Kitano, Shintaro Hayashi, Toshiyuki Kishimoto, Kensuke Kosugi, Satoru Sakurai
概要 This study examined how different pulse oximeter application sites affect blood oxygen saturation (SpO2) measured amount of time and the effects of different pulse oximeter application sites during bodily movements. Thetime from airway obstruction to SpO2 recovery was shortest in the forehead sensor. The greatest error was ear sensor. In dental treatment under intravenous sedation, a pulse oximeter attached to forehead consider most resistant to the effects of bodily movement. However, we consider that the finger site is the best site for recognizing hypoxic states.